Titanium Based Dry Electrodes for Biostimulation and Data Acquisition

Distributed Computing and Artificial Intelligence, 19th International Conference(2022)

引用 0|浏览8
暂无评分
摘要
Wet electrodes rely on conductive electrolyte gel for proper performance, usually presenting high setup time and being disposable or requiring time-consuming cleaning methods. Skin irritation and signal quality deterioration are some of the problems that may occur with these electrodes. Therefore, to overcome these drawbacks, 3D printed bases using Fused Deposition Modelling (FDM) with Polylactic acid (PLA), Polyurethane (PU) and Cellulose filaments were functionalized with titanium (Ti) and titanium-nitride (TiN) thin films and dopped with copper (Cu). The electrodes, implemented using different diameters, were used to record electromyography (EMG) signals proceeded by biostimulation sessions to access their electrical and mechanical characteristics and compare them to that of commercial AgCl/Ag electrodes. The results show that TiN and TiNCu0.45 dry electrodes present results similar to wet AgCl/Ag electrodes for data acquisition. To be comparable to usual carbon electrodes for muscle stimulation, they need some conductivity improvements to lower the necessary voltage. Besides, the endurance of the thin films must be enhanced as well as the adhesion to the polymer.
更多
查看译文
关键词
Dry electrodes, Signal acquisition, Biostimulation, sEMG, Thin film