Screw Pullout Strength After Pedicle Screw Reposition: A Finite Element Analysis

Yu-Xing Ye,Da-Geng Huang,Ding-Jun Hao, Jia-Yuan Liu,Jia-Jia Ji, Jin-Niu Guo

Spine(2023)

Cited 0|Views9
No score
Abstract
Research design.Finite element analysis based on computed tomography images from the lumbar spine.Objective.Determined the pullout strength of unsatisfactorily placed screws and repositioned screws after unsatisfactory place in lumbar spine surgery.Background.Pedicle screws are widely used to stabilize the spinal vertebral body. Unsatisfactory screws could lead to surgical complications, and may need to be repositioned. Screw removal and reposition, however, may decrease pullout strength.Methods.We conducted a three-dimensional finite element analysis based on high-resolution computed tomography images from a 39-year-old healthy woman. Pullout strength was determined with the screw placed in different orientations at the same entry point (as selected by the Magerl method), as well as after removal and reposition. The material properties of the vertebral body and the screw were simulated by using grayscale values and verified data, respectively. A load along the screw axis was applied to the end of the screw to simulate the pullout.Results.The pullout strength was 1840.0 N with the Magerl method. For unsatisfactorily placed screws, the pullout strength was 1500.8 N at 20% overlap, 1609.6 N at 40% overlap, 1628.9 N at 60% overlap, and 1734.7 N at 80% overlap with the hypothetical screw path of the Magerl method. For repositioned screws, the pullout strength was 1763.6 N, with 20% overlap, 1728.3 N at 40% overlap, 1544.0 N at 60% overlap, and 1491.1 N at 80% overlap, with the original path. Comparison of repositioned screw with unsatisfactorily placed screw showed 14.04% decrease in pullout strength at 80% overlap, 5.21% decrease at 60% overlap, 7.37% increase at 40% overlap, and 17.51% increase at 20% overlap, with the screw path of the Magerl method.Conclusions.Removal and reposition increased the pullout strength at 20% and 40% overlap, but decreased the pullout strength at 60% and 80% overlap. For clinical translation, we recommend removal and reposition of the screw when the overlap is in the range of 20% to 40% or less. In vitro specimen studies are needed to verify these preliminary findings.
More
Translated text
Key words
pedicle screw reposition,strength,pull-out
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined