Analysis of urine by MIP-OES: challenges and strategies to correct matrix effects.

Analytical methods : advancing methods and applications(2023)

引用 2|浏览8
暂无评分
摘要
This work investigated the potential of microwave-induced plasma optical emission spectrometry (MIP-OES) for urine analysis using a complex matrix containing carbon and high concentrations of easily ionizable elements (EIEs). The goals were to study interferences originating from the urine matrix for 14 analytes with total energies varying from 1.85 to 12.07 eV, along with strategies to correct matrix effects and compare the results with those reported in the literature using inductively coupled plasma optical emission spectrometry (ICP-OES). It was found that the urine matrix caused suppression of the signals for some elements and increased them for others. Therefore, an internal standardization calibration method and three levels of dilution, , 2-, 20-, and 200-fold, were applied as strategies to correct non-spectral interferences. Also, Ga, Ge, Pd, Rh, Sc and Y and four molecular species present in the nitrogen plasma (, CN, N, N, and OH) were investigated as potential internal standards (ISs). The accuracy and precision were evaluated by addition and recovery experiments and best results were obtained using ISs Ge, Rh and Sc for 20-fold dilution and N for 200-fold dilution. The LODs ranged from 0.33 to 329 μg L and deviations were lower than 11%. The combined use of these strategies led to successful urine analysis for a spiked sample by MIP-OES.
更多
查看译文
关键词
urine,mip-oes