Correlation between osmolarity measurements using the TearLab? and I-Pen? systems in subjects with a high body mass index
ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE(2022)
摘要
Background. Osmolarity is used to detect symptoms of dry eye disease (DED) and can be measured using TearLab (TM) and I-Pen (R) systems. Objectives. To investigate the correlation between osmolarity measurements using the TearLab (TM) and I-Pen (R) systems in subjects with a high body mass index (BMI). Materials and methods. Thirty male subjects with a high BMI (27-48 kg/m(2); 23.3 +/- 2.1 years old) participated in this study. The control group consisting of 30 healthy males (24.9 kg/m(2); 22.9 +/- 2.1 years old) was also enrolled. Osmolarity measurements were recorded from the right eye using the TearLab (TM) and I-Pen (R) systems, and interviews were conducted to determine ocular surface disease index (OSDI) scores. Results. The OSDI (p = 0.042), TearLab (TM) (p < 0.001) and I-Pen (R) (p < 0.001) scores were significantly higher in the study group than in the control group. In the study group, OSDI scores ranged from 2 to 16 (median 8.0, interquartile range (IQR) 6.8), while it was from 0 to 10 (median 6.3, IQR 4.1) in the control group. The TearLab (TM) osmolarity scores were in the range of 278-309 mOsm/L in the study group, whereas the I-Pen (R) osmolarity measurements were in the range of 294-336 mOsm/L in the study group, compared with 263-304 mOsm/L and 278-317 mOsm/L in the control group, respectively. In the study group, there was a strong correlation between the TearLab (TM) and I-Pen (R) osmolarity scores (r = 0.934; p = 0.001). In addition, strong correlations were found between the BMI and both TearLabT (r = 0.736; p = 0.001) and I-Pen (R) (r = 0.707; p = 0.001) scores, as well as between the OSDI scores and both TearLab (TM) (r = 0.731; p = 0.001) and I-Pen (R) measurements (r = 0.666; p = 0.001). Conclusions. Osmolarity measurements using the I- Pen (R) system were significantly higher than those recorded using the TearLab (TM) system in subjects with a high BMI. The I-Pen (R) measurements showed large variations in osmolarity scores and were highly unreliable in correctly identifying normal eyes compared to the TearLab (TM) system. Also, a strong correlation was found between the osmolarity scores obtained from the TearLab (TM) and I-Pen (R) systems.
更多查看译文
关键词
tear film,dry eye,tear osmolarity,high body mass index,TearLab (TM) system,I-Pen (R) system
AI 理解论文
溯源树
样例
![](https://originalfileserver.aminer.cn/sys/aminer/pubs/mrt_preview.jpeg)
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要