Analysis of Polyethylene-Related Revisions After Total Ankle Replacements Reported in US Food and Drug Administration Medical Device Adverse Event Database

Hongying Jiang, Leihong Wu, Per-Henrik Randsborg, Jennifer Houck, Limin Sun, Marisa Marine, Megan Chow, Joseph Peluso, Raquel Peat

Foot & ankle international(2023)

Cited 1|Views10
No score
Abstract
Background: There are 2 general types of total ankle replacement (TAR) designs with respect to the polyethylene insert, mobile-bearing (MB) and fixed-bearing (FB) TARs. The aim of this study is to compare polyethylene-related adverse events (AEs), particularly revisions, reported for MB TARs and FB TARs using the US Food and Drug Administration's (FDA's) Manufacturer and User Facility Device Experience (MAUDE) database. Methods: A text mining method was applied to the medical device reporting (MDR) in the MAUDE database from 1991 to 2020, followed by manual reviews to identify, characterize, and describe all polyethylene-related AEs, including revisions, reported for MB and FB TARs. Results: We found 1841 MDRs for MB (STAR Ankle only) and 1273 MDRs for 40+ FB TARs approved/cleared by the FDA. For the MB design, 33% (606/1841) of the AEs reported related to the polyethylene component, compared to 24% (291/1273) of the AEs reported for FB designs. Polyethylene fractures were reported in 11.3% (208/1841) for the MB designs compared to 0.2% (2/1273) for the FB designs. Half of the polyethylene-related revisions occurred within an average of 4.1 years after implantation for the MB design compared within an average of 5.2 years for FB designs. Conclusion: Analysis of this database revealed a higher proportion of reported polyethylene fractures and greater need for earlier revisions for polyethylene-related issues with use of the primary MB design in the database as compared with FB TAR designs. Further study of device-related complications with more recent designs for both MB and FB ankle replacement components are needed to improve the outcomes of total ankle replacement.
More
Translated text
Key words
total ankle replacement,TAR,adverse event,revision,polyethylene-related adverse events,polyethylene fractures,polyethylene wear,FDA,FDA Manufacturer and User Facility Device Experience,MAUDE,MDR,ultrahigh molecular weight polyethylene,UHMWPE
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined