REACT DX registry: Real world REACTion to atrial high rate episodes detected in implantable cardioverter-defibrillator recipients with a DX lead

Technology and health care : official journal of the European Society for Engineering and Medicine(2023)

引用 0|浏览15
暂无评分
摘要
BACKGROUND: Atrial fibrillation (AF) is associated with significant morbidity and is predicted by atrial high rate events. The early detection of AF is paramount to timely interventions to reduce the morbidity of AF. The DX ICD system combined with Home Monitoring R allows for continuous atrial rhythm monitoring without the need for a dedicated atrial lead. OBJECTIVE: To establish the reaction to and timing of reactions to the detection of atrial high rate episodes (AHRE). METHODS: A prospective cohort of DX ICD systems was followed up and the response to AHREs was collected and evaluated. RESULTS: A total of 234 patients were enrolled; an AHRE > 6 min was detected in 13.7% of patients (n = 32) within a mean follow-up duration of 16 months. A high rate of oral anticoagulation (OAC) prescription was seen with the detection of AHREs in patients with a not-low risk CHA 2DS 2 -VASc score. There was a delay in this prescription highlighting the potential to improve the timeliness of patient care in this group of patients. CONCLUSIONS: The DX ICD system provides rapid and ongoing atrial rhythm monitoring such that physicians are rapidly aware of AHRE without the need for a dedicated atrial lead, but local protocols are needed to improve the response time of anti-coagulation prescription.
更多
查看译文
关键词
DX technology,home monitoring,atrial high rate events,atrial fibrillation,oral anticoagulation