Chrome Extension
WeChat Mini Program
Use on ChatGLM

A novel rotation method with variable-angle anterior probe for correcting the depth of the kidney to optimize renal dynamic imaging

EJNMMI physics(2022)

Cited 0|Views6
No score
Abstract
Purpose We aimed to investigate the effect and significance of the rotation method with variable-angle anterior probe corrected for the depth of two kidneys on the determination of glomerular filtration rate (GFR) in total and single kidneys by the renal dynamic imaging Gates method. Methods Seventy-two patients who underwent dynamic renal imaging by the rotation method and abdominal CT in our hospital were collected in the present study. CT scanning, rotation method, Tonnesen's formula, and Li-Qian's formula were compared in terms of the depth of two kidneys, the depth difference between the two kidneys, and the total renal and single GFR obtained by substituting the renal depth values into Gates' formula. Results ① The depth of kidneys and GFR: Compared to CT, Tonnesen's formula significantly underestimated the depth of both kidneys and the total and single renal GFR ( P < 0.05). No significant differences were found in the depth of both kidneys and the total and single renal GFR between Li-Qian's formula and the rotation method ( P > 0.05), with a strong agreement and with the least bias in the values measured by the rotation method. ②Renal depth difference: Compared to CT, Tonnesen's formula and Li-Qian's formula underestimated the difference in depth between the two kidneys ( P < 0.05). None of the differences were statistically significant based on the rotation method ( P > 0.05). The depth difference was positively correlated with the resulting changes in single renal function (|R(CT)-R(Li-Qian)|) and (|R(Rotation)-R(Li-Qian)|) ( r = 0.881, 0.641, P < 0.001). As the depth difference increased, Li-Qian's formula could not visualize changes in single renal function accurately. In contrast, the accuracy of the rotation method in assessing single renal function remains unaffected. Conclusion The rotation method obtains an accurate depth and depth difference between the two kidneys without additional CT radiation, enhancing the accuracy of the Gates method for determining total and single renal GFR. Trial registration Medical Ethics Committee of First Hospital of Shanxi Medical University, 2021BAL0146. Registered 12 January 2021.
More
Translated text
Key words
Renal depth,Renal dynamic imaging,Rotation method
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined