Quantifying 6D tumor motion and calculating PTV margins during liver stereotactic radiotherapy with fiducial tracking

FRONTIERS IN ONCOLOGY(2022)

引用 0|浏览4
暂无评分
摘要
Objective: Our study aims to estimate intra-fraction six-dimensional (6D) tumor motion with rotational correction and the related correlations between motions of different degrees of freedom (DoF), as well as quantify sufficient anisotropic clinical target volume (CTV) to planning target volume (PTV) margins during stereotactic body radiotherapy (SBRT) of liver cancer with fiducial tracking technique. Methods: A cohort of 12 patients who were implanted with 3 or 4 golden markers were included in this study, and 495 orthogonal kilovoltage (kV) pairs of images acquired during the first fraction were used to extract the spacial position of each golden marker. Translational and rotational motions of tumor were calculated based on the marker coordinates by using an iterative closest point (ICP) algorithm. Moreover, the Pearson product-moment correlation coefficients (r) were applied to quantify the correlations between motions with different degrees of freedom (DoFs). The population mean displacement ( M P over bar ), systematic error (sigma) and random error (sigma) were obtained to calculate PTV margins based on published recipes. Results: The mean translational variability of tumors were 0.56, 1.24 and 3.38 mm in the left-right (LR, X), anterior-posterior (AP, Y), and superior-inferior (SI, Z) directions, respectively. The average rotational angles theta ( X ) , theta ( Y ) and theta ( Z ) around the three coordinate axes were 0.88, 1.24 and 1.12, respectively. (|r|> 0.4) was obtainted between Y -Z , Y - theta ( Z ) , Z -theta ( Z ) and theta ( X ) - theta ( Y ) . The PTV margins calculated based on 13 published recipes in X, Y, and Z directions were 1.08, 2.26 and 5.42 mm, and the 95% confidence interval (CI) of them were (0.88,1.28), (1.99,2.53) and (4.78,6.05), respectively. Conclusions: The maximum translational motion was in SI direction, and the largest correlation coefficient of Y-Z was obtained. We recommend margins of 2, 3 and 7 mm in LR, AP and SI directions, respectively.
更多
查看译文
关键词
liver SBRT, CyberKnife, fiducial tracking, tumor motion, PTV margin
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要