Design and Haemodynamic Analysis of a Novel Anchoring System for Central Venous Pressure Measurement.

Sensors(2022)

Cited 0|Views16
No score
Abstract
This study successfully demonstrated the design and deployment of an inferior vena cava anchoring system in a bench test model. The 10-strut anchor is an optimal design as compared with the two other 5-strut designs; however, substantial in-vivo experiments are required to validate the safety and accuracy of such implants. The CFD simulation enabled better understanding of the haemodynamic parameters and any disturbances in the blood flow due to the presence of the anchor. The ability to place a sensor technology in the vena cava could provide a simple and minimally invasive approach for heart failure patients.
More
Translated text
Key words
anchoring system,bench testing,central venous pressure,computational fluid dynamics,haemodynamic analysis,heart failure,inferior vena cava,remote patient management,wireless pressure sensor