In vitro evaluation of how the presence of the stent retriever and microcatheter influences aspiration parameters in thrombectomy according to their position inside the aspiration catheter.

Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences(2022)

Cited 0|Views7
No score
Abstract
In our experimental model, the position of the stent retriever and microcatheter within the aspirator catheter affected the flow rate, but not the aspiration force. Negative pressure generated within the aspiration catheter appeared to determine a partial collapse of the distal segment that resulted in a less effective aspiration force than the theoretical aspiration force level.
More
Translated text
Key words
stroke, aspiration catheter, stentriever, microcatheter