Chrome Extension
WeChat Mini Program
Use on ChatGLM

Validation of a Miniaturized Handheld Arterial Pressure Monitor for Guiding Full and Partial REBOA Use During Resuscitation

EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY(2023)

Cited 0|Views13
No score
Abstract
Resuscitative endovascular balloon occlusion of the aorta (REBOA) is a well-validated method for the control of noncompressible truncal hemorrhage. In lower resource or battlefield settings, the need for arterial line setup and monitoring is problematic and potentially prohibitive. We sought to evaluate the accuracy and precision of a miniaturized portable device (Centurion COMPASS®) versus standard arterial pressure monitoring using standard ER-REBOA and partial REBOA (pREBOA) as a high-fidelity and space-/time-conserving alternative. A total of 40 swine underwent a four-phase validation/precision study (each phase using five ER-REBOAs and five pREBOAs). Phases I/II evaluated accuracy with full and pREBOA in uninjured animals. Phases III/IV duplicated the previous phases but in a severe hemorrhagic shock model. Carotid and femoral pressures were monitored with both intra-arterial pressure systems and the COMPASS® device. The vascular flow was measured by aortic flow probes. Correlation and Bland–Altman analysis were performed. There was a strong correlation in accuracy testing of proximal and distal COMPASS® devices compared to standard intra-arterial pressure monitoring (r = 0.94, 0.8; p < 0.005) as well as during precision testing (r = 0.98, 0.89 p < 0.005) in the uninjured phases. Similar accuracy and reliability were demonstrated in hemorrhagic shock, with a strong correlation for the proximal and distal COMPASS® devices (r = 0.98, 0.97; p < 0.005), as well as during precision testing (r = 0.99, 0.95; p < 0.005) in both full and pREBOA scenarios. Bland–Altman analysis showed extremely low bias between the COMPASS® and arterial line for both proximal (bias = 1.9) and distal (bias = 0.8) pressure measurements. The COMPASS® provides accurate and precise pressure measurements during standard and partial REBOA in both uninjured and shock conditions. This device may help extend and enhance capability in any low-resource/battlefield settings, or even eliminate the need for standard intra-arterial invasive pressure monitoring and external setup.
More
Translated text
Key words
REBOA,Partial REBOA,Hemodynamic monitoring,Arterial pressure,Military,Prehospital resuscitation
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined