An Advanced Calibration Method for Probe Leakage Correction in On-Wafer Test Systems

IEEE Transactions on Microwave Theory and Techniques(2023)

引用 3|浏览14
暂无评分
摘要
This article presents an advanced calibration method for solving the error terms due to probe-probe leakage in an on-wafer test system. A new 12-term error model for the on-wafer test system including vector network analyzer (VNA), frequency extenders, cables/waveguides, probes, probe contact pads and probe-probe leakage is introduced. A two-step calibration process and an algorithm with four on-chip calibration standards including one undefined Thru, two pairs of undefined symmetrical Reflects such as Open-Open and Short-Short pairs, and a pair of known Match loads has been developed. In addition, an improved circuit model for the Match load is proposed for enhanced accuracy. The calibration method has been tested on a mismatched attenuator for the frequency range between 0.2 and 110 GHz, and the results are compared with numerical simulation and existing calibration methods. It is shown that the attenuator’s $\vert S_{11}\vert $ is more consecutive and $\vert S_{21}\vert $ has been improved by up-to 1.7 dB. It is evident that the proposed calibration method has a simpler calibration process and less stringent requirements on calibration standards which are key for on-wafer system calibration at millimeter-wave and terahertz frequencies. More importantly, the new calibration method is more suitable for measurements in which DUTs have variable lengths.
更多
查看译文
关键词
Calibration,error model,load circuit model,on-wafer scattering parameter,probe leakage