Effect of implant length variations on stress shielding in proximal humeral replacement after tumor excision under torsion: Finite element study.

Journal of orthopaedic research : official publication of the Orthopaedic Research Society(2022)

Cited 0|Views9
No score
Abstract
The proximal humerus is the most common site of occurrence of primary bone tumors in the upper limb. Endoprosthetic replacement is deemed as the preferred reconstructive option following primary resection of bone tumors. However, it has been also associated with complications such as stress shielding and aseptic loosening compromising prosthetic survival. Our objective was to conduct a finite element (FE) study to investigate the effect of varying endoprosthesis length on bone stresses as well as to quantify the extent of stress shielding across the bone length (BL) in a humerus-prosthesis assembly for proximal humeral replacement after tumor excision thereby allowing us to identify the optimal implant length with best biomechanical performance. FE models of the intact humerus and humerus-prosthesis assemblies were established where they were loaded at the elbow joint under torsion with the glenohumeral joint fixed to represent twisting. After dividing the bone into individual slices consisting of 5% BL, the maximum cortical and cancellous principal, von Mises and shear bone stresses were calculated. To measure the level of stress shielding, the percentage stress change from the intact state was evaluated across each slice. Similar stress patterns were observed between the intact state and shorter endoprosthesis compared to the longer endoprostheses. Our findings illustrated the possibility of stress shielding occurring under torsional forces with its effect increasing with implant lengthening. To conclude, we believe that using a shorter prosthesis may substantially diminish the risk of potential implant failure due to stress shielding.
More
Translated text
Key words
endoprosthetic reconstruction,finite element analysis,humeral tumors,implant lengths,stress shielding
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined