Inadvertent QRS prolongation by an optimization device-based algorithm in patients with cardiac resynchronization therapy.

PloS one(2022)

Cited 0|Views4
No score
Abstract
QuickOpt algorithm systematically favours a left-preexcitation VVD which translates into a significant prolongation of the QRSd compared to synchronous biventricular pacing. There is no reason to believe that a manipulation that systematically widens QRSd should be considered to optimize physiology. Device-based CRT optimization algorithms should undergo systematic mechanistic pre-clinical evaluation in various scenarios before they are tested in large clinical studies.