谷歌Chrome浏览器插件
订阅小程序
在清言上使用

Microwave Digestion and ICP-MS Determination of Major and Trace Elements in Waste Sm-Co Magnets

METALS(2022)

引用 1|浏览4
暂无评分
摘要
In this article, inductively coupled plasma mass-spectrometry (ICP-MS) and inductively coupled plasma optical-emission spectrometry (ICP-OES) were used for the development of an analytical procedure for analysis of the waste of Sm-Co magnets. Experimental parameters related to microwave digestion processes and acid concentrations were optimized. Microwave digestion was carried out in mixtures of HF, HCl, HNO3 and H2SO4. The complete dissolution of the samples occurred in the system: 10 mL H2O, 2 mL HNO3, 10 mL HCl and 1 mL H2SO4. The dependence of the matrix effect on the ICP-MS analysis of waste Sm-Co magnets was studied and optimal instrumental parameters were investigated (nebulizer gas flow, sampling depth and potential at the extractor lens). The optimal conditions were a nebulizer gas flow of 0.85-0.90 L/min, a sampling depth of 101, potential at the extractor lens of -400 V and a sample flow rate of 50 rpm. A recovery test and inter-method experiments were performed to verify the accuracy of the proposed method.
更多
查看译文
关键词
rare earth elements, Sm-Co magnets, ICP-MS, ICP-OES, microwave digestion