Verification of dose distribution in high dose-rate brachytherapy for cervical cancer using a normoxic N-vinylpyrrolidone polymer gel dosimeter

Yusuke Watanabe, Takuya Maeyama, Shinya Mizukami, Hidenobu Tachibana, Tsuyoshi Terazaki, Hideyuki Takei, Hiroshi Muraishi, Tsutomu Gomi, Shin-ichiro Hayashi

Journal of radiation research(2022)

引用 2|浏览10
暂无评分
摘要
The polymer gel dosimeter has been proposed for use as a 3D dosimeter for complex dose distribution measurement of high dose-rate (HDR) brachytherapy. However, various shapes of catheter/applicator for sealed radioactive source transport used in clinical cases must be placed in the gel sample. The absorbed dose readout for the magnetic resonance (MR)-based polymer gel dosimeters requires calibration data for the dose-transverse relaxation rate (R-2) response. In this study, we evaluated in detail the dose uncertainty and dose resolution of three calibration methods, the multisample and distance methods using the Ir-192 source and the linear accelerator (linac) method using 6MVX-rays. The use of Ir-192 sources increases dose uncertainty with steep dose gradients. We clarified that the uniformly irradiated gel sample improved the signal-to-noise ratio (SNR) due to the large slice thickness of MR images and could acquire an accurate calibration curve using the linac method. The curved tandem and ovoid applicator used for intracavitary irradiation of HDR brachytherapy for cervical cancer were reproduced with a glass tube to verify the dose distribution. The results of comparison with the treatment planning system (TPS) calculation by gamma analysis on the 3%/2 mm criterion were in good agreement with a gamma pass rate of 90%. In addition, the prescription dose could be evaluated accurately. We conclude that it is easy to place catheter/applicator in the polymer gel dosimeters, making them a useful tool for verifying the 3D dose distribution of HDR brachytherapy with accurate calibration methods.
更多
查看译文
关键词
high dose-rate brachytherapy,Iridium-192,polymer gel dosimeter,quality assurance (QA)
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要