Radiation reduction in a modern catheterization laboratory: A single‐center experience

Catheterization and Cardiovascular Interventions(2022)

引用 0|浏览10
暂无评分
摘要
Background Measures were undertaken at the Cleveland Clinic to reduce radiation exposure to patients and personnel working in the catheterization laboratories. We report our experience with these improved systems over a 7-year period in patients undergoing diagnostic catheterization (DC) and percutaneous coronary interventions (PCIs). Methods Patients were categorized into preinitiative (2009-2012) and postinitiative (2013-2019) groups in the DC and PCI cohorts. Propensity score matching was done between the pre- and postinitiative groups for both cohorts based on age, sex, body surface area, total fluoroscopy time, and total acquisition time. The effectiveness of radiation reduction measures was assessed by comparing the total air kerma (K-a,K-r), and fluoroscopy- and acquisition-mode air kerma in patients in the two groups. Results In the DC cohort, there was a significant reduction in K-a,K-r in the postinitiative group in comparison to the preinitiative group (median, 396 vs. 857 mGy; p < 0.001). In the PCI cohort, K-a,K-r in the postinitiative group was 1265 mGy, which was significantly lower than the corresponding values in the preinitiative group (1994 mGy; p < 0.001). We also observed a significant reduction in fluoroscopy- and acquisition-based air kerma rates, and air kerma area product in the postinitiative group in comparison to the preinitiative group in both matched and unmatched DC and PCI cohorts after the institution of radiation reduction measures. Conclusion There was a significant and sustained reduction in radiation exposure to patients in the catheterization laboratory with the implementation of advanced protocols. Similar algorithms can be applied in other laboratories to achieve a similar reduction in radiation exposure.
更多
查看译文
关键词
catheterization, occupational safety, radiation exposure
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要