Automating QA analysis for a six-degree-of-freedom (6DOF) couch using image displacement and an accelerometer sensor

Physica Medica(2022)

引用 1|浏览1
暂无评分
摘要
The purpose of this study is to develop an approach for automating quality assurance (QA) analysis for a six-degree-of-freedom (6DOF) couch using image displacement and an accelerometer sensor. A cubic phantom was fabricated using 3D printing and the accelerometer sensor was embedded in the phantom to measure the couch in the pitch and roll directions. The accuracy and reliability of image displacement and the accelerometer sensor were investigated prior to their practical use for 6DOF couch QA. Image displacement performance had an accuracy and reliability of 0.026 ± 0.026 mm for the translation direction and 0.021 ± 0.016° for the rotation direction. Accelerometer sensor performance had an accuracy and reliability of 0.023 ± 0.018° for pitch rotation and 0.051 ± 0.024° for roll rotation. Automating QA analysis was used to perform 6DOF couch QA, and the couch position errors measured using image displacement were less than 0.99 mm, 0.91 mm, 0.82 mm for the vertical, longitudinal, lateral translation in range between ±20 mm, and 0.07°, 0.23°, and 0.2° for pitch, roll, and yaw rotation in range between ±3° whereas the couch position errors measured using the accelerometer sensor were less than 0.1° and 0.19° for the pitch and roll rotation in range between ±3°.