CLARION2-TRINITY: A Compton-suppressed HPGe and GAGG:Ce-Si-Si array for absolute cross-section measurements with heavy ions

T. J. Gray,J. M. Allmond, D. T. Dowling,M. Febbraro, T. T. King,S. D. Pain,D. W. Stracener,S. Ajayi, J. Aragon, L. Baby, P. Barber,C. Benetti, S. Bhattacharya, R. Boisseau, J. Gibbons, S. L. Tabor, V. Tripathi,C. Wibisono,I. Wiedenhoever,L. Bignell,M. S. M. Gerathy,G. Lane, L. J. McKie,A. J. Mitchell, J. Pope,R. du Rietz,A. E. Stuchbery, K. Vaigneur, T. J. Ruland

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2022)

Cited 1|Views41
No score
Abstract
The design and performance of a new Compton-suppressed HPGe and charged-particle array, CLARION2-TRINITY, are described. The TRINITY charged-particle array is comprised of 64 Cerium-doped Gadolinium Aluminum Gallium Garnet (GAGG:Ce) crystals configured into five rings spanning 7–54 degrees, and two annular silicon detectors that can shadow or extend the angular coverage to backward angles with minimal γ-ray attenuation. GAGG:Ce is a non-hygroscopic, bright, and relatively fast scintillator with a light distribution well matched to SiPMs. Count rates up to 40 kHz per crystal are sustainable. Fundamental characteristics of GAGG:Ce are measured and presented, including light- and heavy-ion particle identification (PID) capability, pulse-height defects, radiation hardness, and emission spectra. The CLARION2 array consists of up to 16 Compton-suppressed HPGe Clover detectors (≈4% efficiency at 1 MeV) configured into four rings (eight HPGe crystal rings) using a non-Archimedean geometry that suppresses back-to-back coincident 511-keV gamma rays. The entire array is instrumented with 100- and 500-MHz (14 bit) waveform digitizers which enable triggerless operation, pulse-shape discrimination, fast timing, and pileup correction. Finally, two examples of experimental data taken during the commissioning of the CLARION2-TRINITY system are given: a PID spectrum from 16O + 18O fusion-evaporation, and PID and Doppler-corrected γ-ray spectra from 48Ti + 12C Coulomb excitation.
More
Translated text
Key words
Charged particle scintillator,Light quenching,GAGG:Ce,Compton-suppressed high purity germanium array
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined