谷歌Chrome浏览器插件
订阅小程序
在清言上使用

Validation of an implant stability measurement device using the percussion response: a clinical research study

BMC Oral Health(2022)

引用 1|浏览4
暂无评分
摘要
Background Several devices have been developed to measure implant-bone stability as an indicator of successful implant treatment; these include Osstell®, which measures the implant stability quotient (ISQ), and the more recent AnyCheck®, which relies on percussion for the implant stability test (IST). These devices make it possible to measure implant stability. However, no studies have compared the performance of AnyCheck® and Osstell® (i.e., IST and ISQ values) in clinical practice. Therefore, this study aimed to determine the correlation between primary and secondary implant stability using the Osstell® and AnyCheck® devices. Methods Ten patients (7 women; age [mean ± standard deviation]: 49.1 ± 13.3 years) with partially edentulous jaws who received a total of 15 implants were included. IST (AnyCheck®) and ISQ (Osstell®) values were measured immediately after implantation and at 1, 2, 3, 4, and 6 weeks post-implantation. Each measurement was performed three times, and the average value was used as the result. The correlation between measurements obtained using the two devices was determined using Spearman's rank correlation coefficient. Results The IST values ranged from 79.1 ± 2.87 to 82.4 ± 2.65. The ISQ values ranged from 76.0 ± 2.8 to 80.2 ± 2.35. Spearman's rank correlation coefficient was r = 0.64 immediately after implantation, r = 0.29 at 1 week, r = 0.68 at 2 weeks, r = 0.53 at 3 weeks, r = 0.68 at 4 weeks, and r = 0.56 at 6 weeks. A positive correlation was found in all cases, except at week 1 when the correlation was weak; the IST and ISQ values decreased the most during the first postoperative week and increased during the second week. The IST values were also slightly higher at all measurement points. Conclusion The ability to assess implant stability without removing the abutment during healing is essential for determining the timing of loading without the risk of bone resorption. The results of this study suggest that AnyCheck® is useful for determining primary and secondary implant stability.
更多
查看译文
关键词
Dental implant, Implant stability quotient (ISQ), Implant stability test (IST), AnyCheck®, Osstell®
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要