An Exhaled Breath Sampler Based on Condensational Growth and Cyclone Centrifugation (BSCC)

AEROSOL AND AIR QUALITY RESEARCH(2022)

引用 0|浏览8
暂无评分
摘要
An exhaled breath sampler based on condensational growth and cyclone centrifugation (BSCC) was developed and evaluated. The BSCC increases the size of exhaled breath aerosols through condensational growth and then collects them as liquid sample via centrifugation. This enables rapid sample collection and eliminates certain pre-treatment steps for pathogenic microorganism analysis. Laboratory-generated aerosols were mixed with saturated water vapor to simulate exhaled breath, and the collection efficiency and the virus infectivity conservation efficiency of the BSCC were evaluated. The collection efficiency of the BSCC was approximately 66.7% for 100 nm aerosols and increased to nearly 100% for 3 mu m aerosols. Besides, the BSCC maintained approximately 93.5% infectivity of atomized model virus aerosol ( Pseudomonas bacteriophage Phi6). When collecting exhaled breath samples from nine volunteers, the average collection rate was 248.7 mu L min(-1). The BSCC achieved superior overall performance (i.e., 60% high collection efficiency and 40% higher infectivity conservation efficiency) compared with RTube, a commercial used exhaled breath sampler, indicating its potential for diagnosis of respiratory infection and measurements of exhaled viral aerosols.
更多
查看译文
关键词
Viral aerosol collection, Exhaled breath, Sampler development, Collection efficiency, Viral infectivity conservation