Wide-range and calibration-free H2S volume fraction measurement based on combination of wavelength modulation and direct absorption spectroscopy with cavity ringdown spectroscopy

ACTA PHYSICA SINICA(2022)

引用 0|浏览3
暂无评分
摘要
Combining cavity ring down spectroscopy (CRDS) and wavelength modulated direct absorption spectroscopy (WM-DAS), a wide range and calibration-free gas concentration detection method is established, which has the advantages of high signal-to-noise ratio of CRDS and fast speed and measurable absolute concentration of WM-DAS. The baseline ring down time (tau(0)) can be calculated by measuring the absorptivity (WM-DAS) and ring down time (CRDS) at the central frequency of the spectral line, without real-time calibration, which greatly improves the speed of CRDS measurement. The measurement results of different H2S concentrations at 6336.617 cm(-1) at room temperature and atmospheric pressure show that the dynamic measurement range of this method can be extended to more than 4 orders of magnitude, the measurement accuracy is improved in comparison with WM-DAS, and the detection limit can reach 1 x 10(-9) in 40 s.
更多
查看译文
关键词
cavity ring down spectroscopy,wavelength modulation-direct absorption spectroscopy,wide range,calibration free,H2S concentration
AI 理解论文