Optimized Sensor for Broadband Dielectrometry of Biological Liquids of Small Volume

2021 51st European Microwave Conference (EuMC)(2022)

引用 0|浏览2
暂无评分
摘要
In this article, the results of a tailored calibration scheme applied to a transmission line sensor for measuring the relative permittivity of biological liquids in a broad microwave frequency range are discussed. The microwave sensor design enables measurements of relative permittivity values of biological samples with a small volume of only 70nl. The relative permittivity is extracted for aqueous solutions of Sodium Chloride (NaCl) and Bovine Serum Albumin (BSA) over a broadband range from 10MHz to 40 GHz at a constant temperature of 21.5 +/- 0.4 degrees C. The measured permittivity of NaCl solution matches closely with simulated and literature data with relative error of Delta epsilon(r)=0.51 % at 20GHz. Further, the broadband relative permittivity results are compared to extracted permittivity values obtained from a resonant measurement method based on a quartz whispering gallery mode resonator at discrete frequencies between 30 GHz and 40 GHz.
更多
查看译文
关键词
Microwave dielectric spectroscopy, sensor, permittivity, biological liquid