[Evaluation of Sysmex XN-3000 for Detection of Schistocyte].

Zhongguo shi yan xue ye xue za zhi(2022)

引用 0|浏览2
暂无评分
摘要
OBJECTIVE:Compared with the method of optical microscopy, to evaluate the accuracy of fragmented red cells(FRC) detection by Sysmex XN-3000. METHODS:A total of 111 samples were collected from patients diagnosed as thrombotic thrombocytopenic purpura, autoimmune disease, hematological disease, malignant tumor and health examination in our hospital from June 2019 to February 2021, including 74 cases in the case group and 37 cases in the healthy control group. All samples were detected by optical microscope and Sysmex XN-3000, respectively. ROC was used to evaluate the detection ability of Sysmex XN-3000 for schistocyte. Bland-Altman method was used to evaluate the consistency of the results of the two methods for detection of schistocyte, and Pearson correlation analysis was conducted for the difference of the results. RESULTS:The area under the ROC curve was 0.890(95% CI: 0.828-0.952, P<0.01). Sysmex XN-3000 count did not quantitatively agree with schistocyte counts by microscopy in the case group(mean of difference:-1.53, 95% limits of agreement: -8.78~5.72). There was a weak positive correlation between platelet count and the difference of analyzer and microscopic results (r=0.32,P<0.05). CONCLUSION:Sysmex XN-3000 can be used as a reference for qualitative determination of schistocyte. However, the sensitivity of Sysmex XN-3000 should be improved. It is still necessary to combine with manual microscopy. The quantitative results are not reliable now and cannot be used as a reference for monitoring the results of schistocyte in clinical patients after treatment.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要