FDA Emergency Use Authorization-Approved Novel Coronavirus Disease 2019, Pressure-Regulated, Mechanical Ventilator Splitter That Enables Differential Compliance Multiplexing

Michael J Paulsen,Yuanjia Zhu,Matthew H Park,Annabel M Imbrie-Moore,Sam Baker, David Walter Edmonston, Tyler Dawson, Evan Ly, Shannon Martin Bell,Nick A Tran,Jinsuh Jung, Jordan Cedarleaf-Pavy, K R Sridhar, Venkat Venkataraman,Y Joseph Woo

ASAIO JOURNAL(2022)

引用 2|浏览12
暂无评分
摘要
Infection with the novel coronavirus, severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), may cause viral pneumonia and acute respiratory distress syndrome (ARDS). Treatment of ARDS often requires mechanical ventilation and may take weeks for resolution. In areas with a large outbreaks, there may be shortages of ventilators available. While rudimentary methods for ventilator splitting have been described, given the range of independent ventilatory settings required for each patient, this solution is suboptimal. Here, we describe a device that can split a ventilator among up to four patients while allowing for individualized settings. The device has been validated in vitro and in vivo.
更多
查看译文
关键词
COVID-19, ventilator splitter, ARDS, respiratory failure, biomechanical engineering