Bi-exponential fitting excluding b=0 data improves the scan-rescan stability of liver IVIM parameter measures and particularly so for the perfusion fraction.

QUANTITATIVE IMAGING IN MEDICINE AND SURGERY(2022)

引用 0|浏览5
暂无评分
摘要
Background:A prerequisite to translating intravoxel incoherent motion (IVIM) imaging into meaningful clinical applications is sufficient scan-rescan reproducibility. This study aims to confirm the hypothesis that IVIM data fitting by not using b=0 images will improve the stability of liver IVIM measurement. Methods:Healthy volunteers' liver IVIM images were prospectively acquired using a 1.5-T magnet or a 3.0 T with 16 b-values. Repeatability study subjects were scanned twice during the same session, resulted in 35 paired scans for 35 subjects (11 men, mean age: 41.82 years, range: 32-60 years; 24 women, mean age: 42.67 years, range: 20-71 years). IVIM analysis was performed with full-fitting and segmented-fitting with a threshold b-value of 60 s/mm2, and fitting started from b=0 s/mm2 or from b=2 s/mm2. Reproducibility study subjects were scanned and then rescanned with an interval of 5-18 days, resulted in 20 paired scans for 11 subjects (4 men, mean age: 26.25 years, range: 25-27 years; 7 women, mean age: 25.57 years, range: 24-27 years). IVIM analysis was performed with segmented-fitting with a threshold b-value of 50 s/mm2, and fitting started from b=0 s/mm2 or from b=3 s/mm2. Results:Fitting without b=0 data generally improved the repeatability and reproducibility for both PF and Dslow, and particularly so for PF. For with b=0 data segmented fitting repeatability, PF had within-subject standard deviation of 0.019, bland-Atman 75% agreement limit of -31.52% to 28.35%, and ICC of 0.647, while these values were 0.009, -20.78% to 16.86%, and 0.837 for without b=0 analysis. Though the repeatability and reproducibility for Dfast generally also improved, they remained suboptimal. Measurement stability was better for repeatability than for reproducibility. Conclusions:Scan-rescan repeatability and reproducibility of liver IVIM parameters can be improved by fitting without b=0 data, which is particularly so for PF.
更多
查看译文
关键词
Diffusion weighted imaging, liver, intravoxel incoherent motion (IVIM), reproducibility, curve fitting
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要