A case of delayed foreshortening and disconnection of pipeline embolization device detected by chronological cone-beam computed tomography fusion image

Acta Neurochirurgica(2022)

引用 2|浏览14
暂无评分
摘要
Objective The morphological changes of the pipeline embolization device (PED; Medtronic, Minneapolis, MN, USA), such as delayed migration or foreshortening, can relate to the incomplete occlusion of aneurysms. Case presentation A 30-year-old man with a giant cavernous carotid artery aneurysm was treated with two PEDs using the overlapping technique. Six months after treatment, follow-up angiography showed morphological changes of the PEDs and residual flow into the aneurysm. Chronological cone-beam computed tomography fusion imaging clearly revealed the dynamic foreshortening of the first PED and the disconnection of both PEDs, so we decided to implant an additional PED. Conclusion This case illustrates that a three-dimensional understanding can be useful for assessing the cause of treatment failure or recurrence.