Evaluation of the Bonebridge BCI 602 active bone conductive implant in adults: efficacy and stability of audiological, surgical, and functional outcomes

EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY(2022)

Cited 8|Views8
No score
Abstract
Purpose (1) To assess the effectiveness and safety of a bone-conduction implant, the Bonebridge BCI 602, in adults with conductive or mixed hearing loss. (2) To investigate whether the Bonebridge BCI 602 is at least as effective as the Bonebridge BCI 601 in such patients. Methods The study group included 42 adults who had either conductive or mixed hearing loss. All patients underwent Bonebridge BCI 602 implant surgery. Before and after implantation, pure-tone audiometry, speech recognition tests (in quiet and noise), and free-field audiometry were performed. Word recognition scores were evaluated using the Polish Monosyllabic Word Test. Speech reception thresholds in noise were assessed using the Polish Sentence Matrix Test. Subjective assessment of benefits was done using the APHAB (Abbreviated Profile of Hearing Aid Benefit) questionnaire. Results The APHAB questionnaire showed that difficulties in hearing decreased after BCI 602 implantation. Both word recognition in quiet and speech reception threshold in noise were significantly better after BCI 602 implantation and remained stable for at least 12 months. A significant advantage of the device is a reduced time for surgery while maintaining safety. In this study, the mean time for BCI 602 implantation was 28.3 min +/- 9.4. Conclusions The second-generation Bonebridge BCI 602 implant is an effective hearing rehabilitation device for patients with conductive or mixed hearing loss. Patient satisfaction and audiological results confirm its efficacy and safety. Its new shape and dimensions allow it to be used in patients previously excluded due to insufficient or difficult anatomical conditions. The new BCI 602 implant is as effective as its predecessor, the BCI 601.
More
Translated text
Key words
Bone conduction,Bonebridge,Bone conduction implant,Hearing loss,Transcutaneous hearing implant,Quality of life,Partial deafness treatment
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined