Frequency of flash glucose monitoring and glucose metrics: real-world observational data from Saudi Arabia

DIABETOLOGY & METABOLIC SYNDROME(2022)

引用 2|浏览1
暂无评分
摘要
Background This real-world data study analyzed glucose metrics from FreeStyle Libre ® flash glucose monitoring in relation to scanning frequency, time in range (TIR) and estimated A1c (eA1c) in Saudi Arabia. Methods Anonymized reader data were analyzed according to scanning frequency quartiles, eA1c categories (<7%,≥7%‒≤9% or>9%) and TIR categories (<50%,≥50%‒≤70% or>70%). Sensors, grouped by reader, were required to have≥120 h of operation. Differences in scanning frequency, eA1c, TIR, time in hypoglycemia and hyperglycemia, and glucose variability (standard deviation [SD] and coefficient of variation [CV]) were analyzed between groups. Results 6097 readers, 35,747 sensors, and 40 million automatic glucose measurements were analyzed. Patients in the highest scanning frequency quartile (Q4, mean 32.0 scans/day) had lower eA1c (8.47%), greater TIR (46.4%) and lower glucose variation (SD 75.0 mg/dL, CV 38.2%) compared to the lowest quartile (Q1, mean 5.2 scans/day; eA1c 9.77%, TIR 32.8%, SD 94.9 mg/dL, CV 41.3%). Lower eA1c and higher TIR were associated with greater scanning frequency, lower glucose variability and less time in hyperglycemia. Conclusions Higher scanning frequency in flash glucose users from Saudi Arabia is associated with lower eA1c, higher TIR, lower glucose variability and less time in hypoglycemia or hyperglycemia.
更多
查看译文
关键词
FreeStyle Libre ®
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要