High Frequency Bipolar Electroporator with Double-Crowbar Circuit for Load-Independent Forming of Nanosecond Pulses

Vitalij Novickij, Gediminas Staigvila, Arunas Murauskas, Nina Rembialkowska, Julita Kulbacka, Jurij Novickij

APPLIED SCIENCES-BASEL(2022)

Cited 6|Views4
No score
Abstract
In this work, a novel electroporation system (electroporator) is presented, which is capable of forming high frequency pulses in a broad range of parameters (65 ns-100 mu s). The electroporator supports voltages up to 3 kV and currents up to 40 A and is based on H-bridge circuit topology. A synchronized double crowbar driving sequence is introduced to generate short nanosecond range pulses independently of the electroporator load. The resultant circuit generates pulses with repetition frequencies up to 5 MHz and supports unipolar, bipolar, and asymmetrical pulse sequences with arbitrary waveforms. The shortest pulse duration step is hardware limited to 33 ns. The electroporator was experimentally tested on the H69AR human lung cancer cell line using 20 kV/cm bipolar and unipolar 100 ns-1 mu s pulses. Based on a YO-PRO-1 permeabilization assay, it was determined that the electroporator is suitable for applied research on electroporation. The system offers high flexibility in experimental design to trigger various electroporation-based phenomena.
More
Translated text
Key words
electroporation,generator,square wave,crowbar,MHz pulses
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined