Iterative metal artifact reduction on a clinical photon counting system-technical possibilities and reconstruction selection for optimal results dependent on the metal scenario

Julian A Anhaus, Simon Schmidt,Philipp Killermann, Andreas Mahnken,Christian Hofmann

PHYSICS IN MEDICINE AND BIOLOGY(2022)

引用 11|浏览2
暂无评分
摘要
Objective. To give an overview about technical possibilities for metal artifact reduction of the first clinical photon-counting CT system and assess optimal reconstruction settings in a phantom study, assessing monoenergetic imaging (VMI) and iterative metal artifact reduction (iMAR). Approach. Scans were performed with 120 kV and Sn140 kV on the first clinical photon-counting detector CT scanner. To quantify artifact reduction, anthropomorphic phantoms (hip, dental, spine, neuro) were assessed, in addition to a tissue characterization phantom (Gammex) to quantify the HU restoration accuracy, all with removable metal inserts. Each setup was reconstructed with and without dedicated iMAR, and VMIs were computed in 10 keV steps from 40 keV (60 keV at Sn140 kV) to 190 keV for all setups (ground truth and metal with and without iMAR). To find the optimal energy, pixel-wise errors were computed in relevant ROIs in water-equivalent tissue around the metal in each phantom setup. To assess HU restoration potential, measurements were performed in the Gammex phantom's inserts. Main results. Large metal objects (hip head) or metal with high atomic numbers (dental and neuro) do not benefit from higher-energetic reconstructions. The hip shaft (large, low atomic number) comprises a lower base artifact level than the head, still without an energetic optimum. Within the spine (short penetration length, low atomic number) an energy optimum could be identified for both spectra (100 keV for 120 kV and 120 keV for Sn140 kV). The Gammex showed best HU restoration at 100 keV for 120 kV and at 110 keV for Sn140 kV. In all cases, additional iMAR reduced the base artifact level. Significance. This study shows that a novel photon-counting CT system has the capability to reduce metal artifacts in metal types with low atomic number and low penetration length by applying VMI. For all other metal types, additional iMAR is required to reduce artifacts.
更多
查看译文
关键词
photon counting CT, metal artifact reduction, virtual monoenergetic imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要