Superconducting magnet designs and MRI accessibility: A review.

NMR in biomedicine(2023)

引用 4|浏览1
暂无评分
摘要
Presently, magnetic resonance imaging (MRI) magnets must deliver excellent magnetic field (B ) uniformity to achieve optimum image quality. Long magnets can satisfy the homogeneity requirements but require considerable superconducting material. These designs result in large, heavy, and costly systems that aggravate as field strength increases. Furthermore, the tight temperature tolerance of niobium titanium magnets adds instability to the system and requires operation at liquid helium temperature. These issues are crucial factors in the disparity of MR density and field strength use across the globe. Low-income settings show reduced access to MRI, especially to high field strengths. This article summarizes the proposed modifications to MRI superconducting magnet design and their impact on accessibility, including compact, reduced liquid helium, and specialty systems. Reducing the amount of superconductor inevitably entails shrinking the magnet size, resulting in higher field inhomogeneity. This work also reviews the state-of-the-art imaging and reconstruction methods to overcome this issue. Finally, we summarize the current and future challenges and opportunities in the design of accessible MRI.
更多
查看译文
关键词
B0 homogeneity,accessible MRI,imaging in inhomogeneous fields,superconducting magnet design