A voltage trade study for the design of capillary electrophoresis instruments for spaceflight.

Electrophoresis(2023)

引用 0|浏览21
暂无评分
摘要
Capillary electrophoresis (CE) systems have undergone extensive development for spaceflight applications. A flight-compatible high voltage power supply and the necessary voltage isolation for other energized components can be large contributors to both the volume and mass of a CE system, especially if typical high voltage levels of 25-30 kV are used. Here, we took advantage of our custom CE hardware to perform a trade study for simultaneous optimization of capillary length, high voltage level, and separation time, without sacrificing method performance. A capillary electrophoresis with capacitively coupled contactless conductivity detection (CE-C D) method recently developed by our group to target inorganic cations and amino acids relevant to astrobiology was used as a test case. The results indicate that a 50 cm long capillary with 15 kV applied voltage (half of that used in the original method) can be used to achieve measurement goals while minimizing instrument size.
更多
查看译文
关键词
capillary electrophoresis,contactless conductivity detection,high-voltage power supply,in situ analysis,spaceflight instruments