Assessing Mechanical Catheter Dysfunction in Peritoneal Dialysis Patients: A Case Control, Proof-of-Concept Study

semanticscholar(2021)

引用 0|浏览9
暂无评分
摘要
Abstract New guidelines on evaluation of peritoneal membrane function recommend ruling out catheter dysfunction when evaluating patients with low ultrafiltration capacity. We introduce the use of a combination of parameters obtained from daily measurements of the cycler software for predicting catheter dysfunction in automated peritoneal dialysis patients. Out of 117 patients treated between 2015 and 2021, all patients with verified catheter dysfunction (n=14) were identified and compared to controls (n=19). We retrieved cycler data for seven days each and tested parameters predictive capability of catheter dysfunction. Total number of alarms/week >7 as single predictive parameter of catheter dislocation identified 85.7% (sensitivity) of patients with dislocated catheter and 31.6% (1-specificity) of control patients. A combination of parameters (number of alarms/week >7, drain time >22 min, ultrafiltration of last fill <150 mL) where at least two of three parameters appeared identified the same proportion of patients with catheter dislocation, but was more accurate in identifying controls (21% false positive). An easily applicable combination of daily cycler readout parameters, also available in remote monitoring platforms can be used as predictor of inadequate catheter function during routine follow-up with potential for earlier diagnosis of this frequent complication in the future.