Externalized conductors and electrical dysfunction in transvenous ventricular leads: Results of the Cardiac Lead Assessment Study

Roger A Freedman,Anne B Curtis, Stephanie M Delgado, Li-Yin Lee

Heart Rhythm O2(2022)

引用 0|浏览3
暂无评分
摘要
A high prevalence of externalized conductors was found in Riata and Riata ST defibrillator leads, with a higher risk of externalization for 8F Riata leads than for 7F Riata ST leads. The 98% reduction in prevalence of externalized conductors in Durata leads compared to Riata/Riata ST leads confirms that the design improvements culminating in Durata leads significantly improved abrasion resistance and durability.
更多
查看译文
关键词
Electrical dysfunction,Externalized conductor,High-voltage lead,Implantable cardioverter-defibrillator,Lead insulation