Pulse Wave Analysis Using the Pressure Recording Analytical Method to Measure Cardiac Output in Pediatric Cardiac Surgery Patients: A Method Comparison Study Using Transesophageal Doppler Echocardiography as Reference Method

ANESTHESIA AND ANALGESIA(2022)

引用 1|浏览3
暂无评分
摘要
BACKGROUND: Cardiac output (CO) is a key determinant of oxygen delivery, but choosing the optimal method to obtain CO in pediatric patients remains challenging. The pressure recording analytical method (PRAM), implemented in the MostCareUp system (Vygon), is an invasive uncalibrated pulse wave analysis (PWA) method to measure CO. The objective of this study is to compare CO measured by PRAM (PRAM-CO; test method) with CO simultaneously measured by transesophageal Doppler echocardiography (TEE-CO; reference method) in pediatric patients. METHODS: In this prospective observational method comparison study, PRAM-CO and TEE-CO were assessed in pediatric elective cardiac surgery patients at 2 time points: after anesthesia induction and after surgery. The study was performed in a German university medical center from March 2019 to March 2020. We included pediatric patients scheduled for elective cardiac surgery with arterial catheter and TEE monitoring. PRAM-CO and TEE-CO were compared using Bland-Altman analysis accounting for repeated measurements per subject, and the percentage error (PE). RESULTS: We included 52 PRAM-CO and TEE-CO measurement pairs of 30 patients in the final analysis. Mean +/- SD TEE-CO was 2.15 +/- 1.31 L/min (range 0.55-6.07 L/min), and mean PRAM-CO was 2.21 +/- 1.38 L/min (range 0.55-5.90 L/min). The mean of the differences between TEE-CO and PRAM-CO was -0.06 +/- 0.38 L/min with 95% limits of agreement (LOA) of 0.69 (95% confidence interval [CI], 0.53-0.82 L/min) to -0.80 L/min (95% CI, -1.00 to -0.57 L/min). The resulting PE was 34% (95% CI, 27%-41%). CONCLUSIONS: With a PE of <45%, PRAM-CO shows clinically acceptable agreement with TEE-CO in hemodynamically stable pediatric patients before and after cardiac surgery. (PWA) by pressure recording analytical method (PRAM-CO) in pediatric patients undergoing elective major cardiac surgery? Findings: In 30 pediatric patients, with a mean +/- SD TEE-CO of 2.15 +/- 1.31 L/min and a mean PRAM-CO of 2.21 +/- 1.38 L/min, Bland-Altman analysis comparing PRAM-CO and TEE-CO shows 95% limits of agreement of 0.69 (95% confidence interval [CI], 0.53-0.82 L/min) to -0.80 L/min (95% CI, -1.00 to -0.57 L/min) and a resulting percentage error of 34% (95% CI, 27%-41%). Meaning: In this study setting, PRAM-CO shows clinically acceptable agreement with simultaneous TEE-CO and PRAM-CO that can be used as an alternative to TEE-CO to assess CO in hemodynamically stable pediatric patients.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要