High-precision background correction and artifact suppression for ultrafast spectroscopy by quasi-simultaneous measurements in a split-sample cell

REVIEW OF SCIENTIFIC INSTRUMENTS(2022)

引用 4|浏览7
暂无评分
摘要
Alternating acquisition of background and sample spectra is often employed in conventional Fourier-transform infrared spectroscopy or ultraviolet-visible spectroscopy for accurate background subtraction. For example, for solvent background correction, typically a spectrum of a cuvette with solvent is measured and subtracted from a spectrum of a cuvette with solvent and solute. Ultrafast spectroscopies, though, come with many peculiarities that make the collection of well-matched, subtractable background and sample spectra challenging. Here, we present a demountable split-sample cell in combination with a modified Lissajous scanner to overcome these challenges. It allows for quasi-simultaneous measurements of background and sample spectra, mitigating the effects of drifts of the setup and maintaining the beam and sample geometry when swapping between background and sample measurements. The cell is moving between subsequent laser shots to refresh the excited sample volume. With less than 45 mu l of solution for 150 mu m optical thickness, sample usage is economical. Cell assembly is a key step and covered in an illustrated protocol.& nbsp;(c) 2022 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).