Prospective Study Assessing Impact of Ethylene Oxide Sterilization on Endoscopic Ultrasound Image Quality.
Clinical gastroenterology and hepatology : the official clinical practice journal of the American Gastroenterological Association(2022)
Abstract
In this prospective study, expert review and phantom-based testing demonstrated minimal differences in image quality between echoendoscopes reprocessed using HLD vs ETO + HLD over 2 years of clinical use. Further studies are warranted to assess the long-term clinical impact of these findings. In the interim, these results support use of ETO sterilization of EUS instruments if deemed clinically necessary.
MoreTranslated text
Key words
Disinfection,Endoscopic Ultrasound,Ethylene Oxide,Sterilization