Comparison of manual and automatic cell count methods for synovial fluid: A prospective study

Journal of Surgery and Medicine(2022)

引用 0|浏览4
暂无评分
摘要
Background/Aim: Cell count measurement methods in the synovial fluid are still a current problem in orthopedic practice. Knowing the cell count in the synovial fluid is important for the assessment of a variety of orthopedic and rheumatologic diseases. We aimed to assess the correlation of WBC and RBC results obtained with a complete blood count analyzer with that obtained by a manual cell count. Methods: The WBC and RBC count in the synovial fluid of 43 patients undergoing gonarthrosis surgery were determined by a Mindray BC-6800 hematology analyzer. The study groups were defined as manual cell count (MC), hemogram mode (HM), and body fluid mode (BFM). All samples were analyzed twice consecutively, and the mean results were calculated. Cell counting was performed using different methods in the same samples and compared statistically. Results: The mean age of the patients was 60.9 years, and there were 17 males (39.5%) and 26 females (60.5%). The WBC and RBC counts in the synovial fluid samples were determined using manual cell count, and the HM and BFM on a Mindray BC-6800 automatic hematology analyzer. WBC counts significantly differed between MC and BFM, and RBC counts significantly differed between HM-MC and HM-BFM (P=0.001, P=0.001, P=0.001, respectively). There was a significant positive correlation between BFM and MC in WBC counts (r=0.633, P
更多
查看译文
关键词
synovial fluid,automatic cell count methods
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要