Biomechanical Test and Assessment on three Internal Fixations on Spine Subaxial Cervical

Chinese Journal of Traditional Medical Traumatology & Orthopedics(2007)

引用 0|浏览0
暂无评分
摘要
Objectives:To evaluate the primary stabilities of 3 different instrumentations in the subaxial cervical spine and to provide a theory basis for the selection internal fixations reasonably in clinical.Methods:Five fresh human cadavers cervical specimens(C3~C7)were loaded nondestructively with 2.0Nm pure moments.A decompression laminectomy instability was created in C4~C5 and the motion at C4~C6 was measured in nexion-extension,lateral bending,and axial rotation.The five specimens were instrumented by the following fixation techniques,respectively,1.the intact state 2.after the C4~C5 decompression laminectomy 3.facet wiring with bone graft 4.lateral mass screws.5.pedicle screws.Data analysis was performed by calculating the ranges of motion,and statistical differences were determined by the test for group comparison.Results:No significant difference was observed in nexion-extension between pedicle screws and lateral mass screws.Pedicle screw showed a significantly higher stabilitv in torsion.No difference was found between facet wiring with bone graft and lateral mass screws in lateral bending,in other load directions the stability was even worse.Conclusions:The major investigation of the current study was the higher stability of pedicle screws over lateral mass fixation and facet wiring with bone graft with respect to primary stability and stability.From the biomechanical point of view the use of pedicle screws in the subaxial cervical spine seems justified.In decompression laminectomy,lateral mass screws could meet the need in clinics.The stability of facet wiring with bone graft was worse.
更多
查看译文
关键词
biomechanical test,internal fixations
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要