The radiobiological effect of using Acuros XB vs anisotropic analytical algorithm on hepatocellular carcinoma stereotactic body radiation therapy.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists(2022)

引用 1|浏览11
暂无评分
摘要
The purpose of this work was to study the radiobiological effect of using Acuros XB (AXB) vs Analytic Anisotropic Algorithm (AAA) on hepatocellular carcinoma (HCC) stereotactic body radiation therapy (SBRT). Seventy SBRT volumetric modulated arc therapy (VMAT) plans for HCC were calculated using AAA and AXB respectively with the same treatment parameters. Published tumor control probability (TCP) and normal tissue complication probability (NTCP) models were used to quantify the effect of dosimetric difference between AAA and AXB on TCP, NTCP and uncomplicated tumor control probability (UTCP). There was an average decrease of 2.5% in 6-month TCP. Normal liver has the largest average decrease in NTCP which was 59.7%. Bowels followed with 26.6% average decrease in NTCP. Duodenum, stomach and esophagus had 10.2%, 5.1%, and 4.3% average decrease in NTCP. There was an average decrease of 1.8% and up to 7.2% in 6-month UTCP. There was an overall decrease in TCP, NTCP, and UTCP for HCC SBRT plans calculated using AXB compared to AAA which could be clinically significant.