Comparison of the Icare ic100 rebound tonometer and the Goldmann applanation tonometer in 1000 eyes

Ophthalmic Research(2020)

引用 8|浏览0
暂无评分
摘要
Background: Measurement of intraocular pressure (IOP) forms a crucial component in the diagnosis and management of glaucoma. Many devices have been developed to measure IOP with Goldmann applanation tonometry (GAT) considered the gold standard for IOP measurement. Objectives: The objective was to compare the IOP measured using Icare ic100 and GAT. Methods: This cross-sectional study measured IOP in 1,000 eyes (500 left and 500 right) using the Icare ic100 and GAT. Central corneal thickness (CCT) was measured using a hand-held pachymeter. IOP measurements were investigated in all eyes, by IOP substrata, by CCT group, by number of topical glaucoma medications, and diagnosis. Results: There was moderate agreement between ic100 and GAT IOP measurements (intraclass correlation coefficient 0.73). Mean IOP was significantly lower when measured by ic100 than by GAT (12.1 vs. 16.2 mm Hg, p < 0.0001). Mean ic100 IOPs were also significantly lower than mean GAT IOPs within each IOP strata ≤12 (7.9 vs. 9.7 mm Hg, p < 0.0001), 13–21 (12.1 vs. 16.6 mm Hg, p < 0.0001), and ≥22 (18.4 vs. 25.2 mm Hg, p < 0.0001) and within each subanalysis. Conclusions: The Icare tonometer consistently under estimated IOP compared to GAT, irrespective of CCT ranges and other subgroup analyses. The mean difference of 4.2 mm Hg can have significant clinical implications, particularly in the management of glaucoma patients.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要