All Fiber-Optic Immunosensors Based on Elliptical Core Helical Intermediate-Period Fiber Grating with Low-Sensitivity to Environmental Disturbances

BIOSENSORS-BASEL(2022)

引用 4|浏览21
暂无评分
摘要
An all fiber-optic immunosensor based on elliptical core helical intermediate-period fiber grating (E-HIPFG) is proposed for the specific detection of human immunoglobulin G (human IgG). E-HIPFGs are all-fiber transducers that do not include any additional coating materials or fiber architectures, simplifying the fabrication process and promising the stability of the E-HIPFG biosensor. For human IgG recognition, the surface of an E-HIPFG is functionalized by goat anti-human IgG. The functionalized E-HIPFG is tested by human IgG solutions with a concentration range of 10-100 mu g/mL and shows a high sensitivity of 0.018 nm/(mu g/mL) and a limit of detection (LOD) of 4.7 mu g/mL. Notably, the functionalized E-HIPFG biosensor is found to be insensitive to environmental disturbances, with a temperature sensitivity of 2.6 pm/degrees C, a strain sensitivity of 1.2 pm/mu epsilon, and a torsion sensitivity of -23.566 nm/(rad/mm). The results demonstrate the considerable properties of the immunosensor, with high resistance to environmental perturbations, indicating significant potential for applications in mobile biosensors and compact devices.
更多
查看译文
关键词
elliptical core helical intermediate-period fiber grating, immunosensor, human immunoglobulin, all fiber-optic sensor, insensitivity to environmental disturbances