Dental-CT: Bildqualität und Strahlenexposition in Abhängigkeit von den Scanparametern

RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren(2008)

Cited 12|Views1
No score
Abstract
To develop a scan protocol for dental-CT which guarantees good image quality at the lowest possible radiation dose.In an experimental investigation Dental-CT (HSA, GE, Milwaukee, USA) of the mandible of two human skeletons positioned in a water tank were performed in order to define the most advantageous scan protocol. Tube currents ranged from 40 to 200 mA and the scan technique was modified (axial mode or helical mode with pitches of 1 to 3 and corresponding increments of 0.4 to 1.0 mm). 39 patients underwent a dental-CT with decreased current (80 mA) in the helical scan mode (pitch 2, slice thickness 1 mm). Dose measurements were performed for two different scan protocols (A: axial, 130 mAs, B: helical, 80 mA, pitch 2).The preliminary investigations of image quality showed only a minor effect of the applied current. For the helical scan mode, pitches of more than 2 impaired image quality. A low increment had no advantages. There were no disadvantages in clinical practice using protocol B with decreased tube current. Absorbed radiation dose of dental CT performed with protocol B was decreased to one third in comparison to protocol A.A scan protocol with a low tube current (e.g., 80 mA, for a rotation time of 1 s) and a helical scan mode (e.g., for a slice thickness of 1 mm with a pitch of 2 and an increment of 1 mm) is recommended for performing dental-CT.
More
Translated text
Key words
bildqualität und strahlenexposition
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined