The Study of Decrease Artifact when Scanning Head and Neck PET/CT(Decrease Radiation Dose of Patient)

Journal of the Korea Safety Management and Science(2016)

Cited 0|Views0
No score
Abstract
Beam hardening artifact can be caused by metal material when performing PET exam. Therefore, we studied a solution decreasing artifact caused by metallic dental implant. The higher voltage, the lesser artifact in CT exam. But Higher voltage dosen't affect PET exam. The thicker silicon the lesser artifact in CT and PET exam. Both methods make less artifact in CT and PET exam. But considering safety of patient, the way of using silicon is better.