An evaluation of the use of EBT-XD film for SRS/SBRT commissioning of a 1.5 Tesla MR-Linac system

Physica Medica(2022)

引用 0|浏览2
暂无评分
摘要
•ERE has significant impacts on film dosimetry in a 1.5T MR-Linac.•Applying water droplet technique to film dosimetry minimizes the ERE impact.•EBT-XD films provide accurate small field dosimetry under ERE.•Penumbra uncertainties can affect the SRS feasibility with a 1.5T MR-Linac.
更多
查看译文
关键词
MR-Linac,SRS,SBRT,QA,Film dosimetry
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要