Deep learning reconstruction improves radiomics feature stability and discriminative power in abdominal CT imaging: a phantom study

European Radiology(2022)

引用 17|浏览4
暂无评分
摘要
Objectives To compare image quality of deep learning reconstruction (AiCE) for radiomics feature extraction with filtered back projection (FBP), hybrid iterative reconstruction (AIDR 3D), and model-based iterative reconstruction (FIRST). Methods Effects of image reconstruction on radiomics features were investigated using a phantom that realistically mimicked a 65-year-old patient’s abdomen with hepatic metastases. The phantom was scanned at 18 doses from 0.2 to 4 mGy, with 20 repeated scans per dose. Images were reconstructed with FBP, AIDR 3D, FIRST, and AiCE. Ninety-three radiomics features were extracted from 24 regions of interest, which were evenly distributed across three tissue classes: normal liver, metastatic core, and metastatic rim. Features were analyzed in terms of their consistent characterization of tissues within the same image (intraclass correlation coefficient ≥ 0.75), discriminative power (Kruskal-Wallis test p value < 0.05), and repeatability (overall concordance correlation coefficient ≥ 0.75). Results The median fraction of consistent features across all doses was 6%, 8%, 6%, and 22% with FBP, AIDR 3D, FIRST, and AiCE, respectively. Adequate discriminative power was achieved by 48%, 82%, 84%, and 92% of features, and 52%, 20%, 17%, and 39% of features were repeatable, respectively. Only 5% of features combined consistency, discriminative power, and repeatability with FBP, AIDR 3D, and FIRST versus 13% with AiCE at doses above 1 mGy and 17% at doses ≥ 3 mGy. AiCE was the only reconstruction technique that enabled extraction of higher-order features. Conclusions AiCE more than doubled the yield of radiomics features at doses typically used clinically. Inconsistent tissue characterization within CT images contributes significantly to the poor stability of radiomics features. Key Points • Image quality of CT images reconstructed with filtered back projection and iterative methods is inadequate for the majority of radiomics features due to inconsistent tissue characterization, low discriminative power, or low repeatability. • Deep learning reconstruction enhances image quality for radiomics and more than doubled the feature yield at doses that are typically used in clinical CT imaging. • Image reconstruction algorithms can optimize image quality for more reliable quantification of tissues in CT images.
更多
查看译文
关键词
Tomography, X-ray computed,Phantoms, imaging,Liver neoplasms,Algorithms,Reproducibility of results
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要