A multi-institutional comparison of dosimetric data for a 0.35 T MR-linac

Ahtesham Ullah Khan, Rajiv Lotey, Larry A. DeWerd, Poonam Yadav

PHYSICS IN MEDICINE AND BIOLOGY(2022)

引用 4|浏览4
暂无评分
摘要
Objective. A comparison of percent depth dose (PDD) curves, lateral beam profiles, output factors (OFs), multileaf collimator (MLC) leakage, and couch transmission factors was performed between ten institutes for a commercial 0.35 T MR-linac. Approach. The measured data was collected during acceptance testing of the MR-linac. The PDD curves were measured for the 3.32 x 3.32 cm(2), 9.96 x 9.96 cm(2), and 27.20 x 24.07 cm(2) field sizes. The lateral beam profiles were acquired for a 27.20 x 24.07 cm(2) field size using an ion chamber array and penumbra was defined as the distance between 80% of the maximum dose and 20% of the maximum dose after normalizing the profiles to the dose at the inflection points. The OFs were measured using solid-state dosimeters, whereas radiochromic films were utilized to measure radiation leakage through the MLC stacks. The relative couch transmission factors were measured for various gantry angles. The variation in the multi-institutional data was quantified using the percent standard deviation metric. Main results. Minimal variations (<1%) were found between the PDD data, except for the build-up region and the deeper regions of the PDD curve. The in-field region of the lateral beam profiles varied <1.5% between different institutions and a small variation (<0.7 mm) in penumbra was observed. A variation of <1% was observed in the OF data for field sizes above 1.66 x 1.66 cm(2), whereas large variations were shown for small-field sizes. The average and maximum MLC leakage was calculated to be <0.3% and <0.6%, which was well below the international electrotechnical commission (IEC) leakage thresholds. The couch transmission was smallest for oblique beams and ranged from 0.83 to 0.87. Significance. The variation in the data was found to be relatively small and the different 0.35 T MR-linacs were concluded to have similar dosimetric characteristics.
更多
查看译文
关键词
acceptance testing,0,35T MR-linac,multi-institutional,PDD,beam profiles,output factors,MLC leakage
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要