Fitness accuracy and retentive forces of milled titanium clasp.

Dental materials journal(2022)

引用 2|浏览4
暂无评分
摘要
Since cast titanium prostheses have many drawbacks, multi-directionally forged titanium grade 2 (MDF) was developed, and the application of the milling process was proposed for improving the titanium clasp. This in vitro study evaluated milled titanium clasps, including MDF titanium. Milling clasps were manufactured with commercially pure (CP) titanium grade 2 (CP 2), grade 4 (CP 4), Ti-6-Al-4V, and MDF. As a control, a CP 2 cast titanium clasp was fabricated in the conventional manner. No porosities and catastrophic failures were observed in the four milled titanium clasps. Fitness accuracy and retentive forces of milled CP 2 and CP 4 tended to be worse, and the milled MDF showed the higher retentive forces (12.45 N) than did cast and milled CP 2 clasps (9.32 N and 4.42 N). Milled titanium clasps can be recommended for longer-term clinical use as compared to cast clasps.
更多
查看译文
关键词
MDF,Milling,Removable partial denture,Retentive force,Titanium clasp
AI 理解论文