Intraoperative hyperspectral imaging (HSI) as a new diagnostic tool for the detection of cartilage degeneration

SCIENTIFIC REPORTS(2022)

引用 0|浏览7
暂无评分
摘要
To investigate, whether hyperspectral imaging (HSI) is able to reliably differentiate between healthy and damaged cartilage tissue. A prospective diagnostic study was performed including 21 patients undergoing open knee surgery. HSI data were acquired during surgery, and the joint surface’s cartilage was assessed according to the ICRS cartilage injury score. The HSI system records light spectra from 500 to 1000 nm and generates several parameters including tissue water index (TWI) and the absorbance at 960 nm and 540 nm. Receiver operating characteristic curves were calculated to assess test parameters for threshold values of HSI. Areas with a cartilage defect ICRS grade ≥ 3 showed a significantly lower TWI ( p = 0.026) and higher values for 540 nm ( p < 0.001). No difference was seen for 960 nm ( p = 0.244). For a threshold of 540 nm > 0.74, a cartilage defect ICRS grade ≥ 3 could be detected with a sensitivity of 0.81 and a specificity of 0.81. TWI was not suitable for cartilage defect detection. HSI can provide reliable parameters to differentiate healthy and damaged cartilage. Our data clearly suggest that the difference in absorbance at 540 nm would be the best parameter to achieve accurate identification of damaged cartilage.
更多
查看译文
关键词
Medical research,Translational research,Science,Humanities and Social Sciences,multidisciplinary